Guide to OPKUD and
OPBOX
Ultrasonic testing units
Software Revision 3.0 / 2003 |
| |
|

|
 |
OPKUD-01/100 is designed specially for
ultrasonic measurements with the need of A-scan pictures as well as
other kinds of measurements which employ mechanical scanning elements or
multiplexed channels (the card is capable of controlling such devices) -
where B or C-scans can be a result. Together with the pulser&receiver
unit OPGUD-01 (placed in a small box), ultrasonic probe and software it
could be used as complete ultrasonic testing device. OPKUD-01/100 is a
short (8-bit) ISA card and can be installed in a standard PC computer.
OPGUD card can be delivered mounted in a box, containing power supply
and battery, that can be used with a portable computer, equipped with a
centronics interface. This device is named as OPBOX. There are two
versions of OPBOX: one version can be used with a normal OPGUD pulser&receiver,
the second one contains additional electronics allowing to control the
special version of pulser&receiver, containing additional amplifiers
and filters. Both card and a box can be supplied with 8 channel
multiplexer, where each channel can send and receive in any sequence (the
same channel can send and receive too). Each channel of this multiplexer
has the same features as OPGUD pulser&receiver.
|
| Technical
data OPKUD Card |
|
A/D converter: |
|
- Resolution: |
8
bit |
|
- Sampling frequency: |
50
or 100MHz 1 |
| Analog
parameters: |
|
- Input channels: |
2
(switched): send & receive and receive |
|
- Input amplifier gain: |
0dB,
6dB, 14dB, 20dB, 26dB, 34dB, 40dB (hardware amplifier on the card)1 additional
33dB in echo mode (amplifier in the pulser & receiver box)total
possible gain: 110dB (with software amplification) |
|
- Input attenuator |
10dB1 |
|
- Minimal sensitivity: |
0.1mV/Div
-1mVpp (with software amplification) |
|
- Maximal sensitivity: |
50mV/Div
- 0.5Vpp |
| Averaging |
1
- 256 |
| Bandwidth: |
0,1
- 25MHz |
| Input
impedance: |
50Ohm,
10pF |
| Trigger: |
internal
(software), external, max 2 kHz |
| Delay
time: |
Post
trigger 256us, measurement accuracy better than 1ns |
| Pulser: |
Step
pulser, 50V -300V in 8 levels (the voltage depends on the used transducer),
20ns edge falling time, separate, match box size |
| Data
buffer: |
256,
512, 1K or 16K1 |
|

Technical description of the OPKUD device
|
One of the most important features
of the card is a precise synchronization between the trigger signal and
the moment when the sampling of the input signal starts. This
synchronization is significantly better than 1ns, that causes, that time
of flight measurements can be made with accuracy better than 0.1ns.
It is particularly important in the case of scanning devices, since it
allows to achieve a very small time skew between different positions (channels).
1ns would correspond to a clock frequency of 1GHz which is much higher
than the actual frequency used, but is achieved with much cheaper
electronics.
|
| OPGUD-01
Pulser&Receiver |
The basic version of our pulser&receiver circuit have the size of a
matchbox and integrated amplifier. It is powered directly from our
oscilloscope card and enable the adjustment of the transmitter voltage
directly from a PC. This device is intended for use with our OPKUD card
or OPBOX.
|

|
| Specifications: |
Pulser: Step pulser, 50V -360V in 8
levels (the voltage depends on the used transducer), 20ns edge falling
time, separate,
Amplifier: 40dB
Size: 40x85x25mm
|
| Features: |
Pulser circuit waveforms: The rising edge
of the Trigger signal (described as Trig Inp) initiates the transducer
charging process which takes about 3us. After this time a transistor
switch which discharges the transducer is turned on (discharge time is
about 20ns, but it can be longer, if the capacitance of the transducer
is too large; the limit is reached, if the transducer is made from the
standard ceramics, is 0.1 mm thick and has 8 mm
diameter).
|
| Comments: |
Because of the very low output impedance
of the device (<1 Ohm) and short discharge time the pulse generated
with thisdevice could be concerned as a real pulse answer for the most
transducers. Transducers with a parallel matching inductance should not
be used, since they do not allow the transducer to be pre-charged (the
inductance causes a short-circuit).
|
| Signal
form |
|

|
|
Ultrasonic Testing Box
OPBOX-01/100 |
| |
|
Price: 1900,- Euro,
together with our standard 10MHz probe - 1999,- Euro. |
The device contains in the box of the
size of the notebook computer the OPKUD Ultrasonic testing card, pulser&receiver,
interface and a separate power supply with own battery and could be
connected to any computer with Centronics interface.
|
The device is well suited for ultrasonic measurements (Flaw detection,
layer thickness measurements, TOFD etc.). It can be used with mechanical
scanning elements or many multiplexed channels (the device is capable of
controlling such devices). It can work in echo mode (with one transducer
sending and receiving) and with two transducers (transmit mode, TOFD
etc.). Software is included (Windows 9x, NT). The device can also work
as standard oscilloscope.
|
 |
|
Technical data |
| A/D
converter: |
| -
Resolution: |
8 bits |
| -
Sampling frequency: |
50 or 100MHz
1 |
| Analog
parameters: |
|
- Input channels: |
2:
send&receive and receive |
|
- Input amplifier gain: |
0dB,
6dB, 14dB, 20dB, 26dB, 34dB, 40dB (hardware amplifier on the card)1
additional 33dB in echo mode (amplifier in the pulser&receiver box)
total possible gain: 110dB (with software amplification) |
|
- Minimal sensitivity: |
0.1mV/Div
-1mVpp (with additional software amplification) |
|
- Maximal sensitivity: |
50mV/Div
- 0.5Vpp |
|
- Averaging |
1
- 256 |
|
- Bandwidth: |
0,1
- 25MHz |
|
- Input impedance: |
50Ohm,
10pF |
| Data
buffer: |
256,
512, 1K or 16K 1 |
| Trigger: |
internal
(software), external, max 2 kHz |
| Delay
time: |
measurement
accuracy better than 1ns |
| Pulser: |
Step
pulser, 50V -300V in 8 levels (the voltage depends on the used transducer),
20ns edge falling time, separate match sized box |
| Interface: |
Parallel
(Centronics DB25 - female) |
| Power
supply: |
220
or 110V, 4 hours work with the internal battery Sealed Lead-Acid |
| Size: |
255x295x50mm |
| Weight: |
2700g |
| |
1 - software
selected |
| Software
features: |
- Two windows: RF signal in the main
window, in the second optional: zoom, FFT, rectified signal, filtered
signal;
- Software amplification, filtering and averaging (allows to work with
very noisy and weak signals);
- Build in DAC and TCG curves;
- Three signal memories, each signal can be stored and compared with any
other signal.
Optional it is possible to make
software versions for the special needs of the user.
|
|

Technical description of the OPBOX device
|
|
Special External Pulser/Receiver with Amplifier |
The Pulser/Receiver is equipped with
additional amplifiers and can be used only with modified version of
OPBOX, containing an interface for controlling it. One amplifier is used
for PE (Pulse-echo - one-transducer) mode and second one for TT (two-transducers
or through transmit ) mode.
Gain range ofboth of amplifiers is 0 - 50 dB controlled with 0.2 dB
step.
|
|

Front view of Pulser/Receiver

|
| Rear
view of Pulser/Receiver |
|

Diagram of connections of all parts of the
system.
The "Diagram"
shows how all parts of the system are connected. "T1" and
"out1" serves in one-transducer mode, connecting additionally
"T2" and "out2" you can work with two-transducers
mode where T1 is sending transducer and T2 is receiving one.
The Pulser&Receiver is featured with switch able pass filter. It is
0.5 to 8 MHz or 0.5 to 15 MHz bandwidth.
Using this version of pulser&receiver the amplification on both
channels (TT and PE) is the same and can be controlled from computer.
|
|
Software Revision 3.0 / 2003 |
In order to use the Ultrasonic Testing
Card OPGUD or OPBOX-01/100 with a described software, the following
equipment is necessary:
Computer System: IBM PC
Display Adapter: SVGA Adapter working in mode 640x480 or higher
(recommended 800x600)
Operations System: Windows 95, 98, ME, XP or Windows NT or 2000.
Interface for OPBOX: Centronics with EPP mode
|
| Main
Panel Overview: |
|

|
| Overview
of Hotkey Selection: |
[F1] RUN / STOP enables / disables
selected acquisition mode
[F2] Help About as…
[F3] Load settings… Load in a previously saved data file and
restore the capture settings as they were when the data was saved
[F4] Save settings… This option creates one file that contains
both the current settings and the current data. The setting saved are
the same in the Save Settings option.
[F5] Print This option will create a hardcopy of the screen into IBM
Graphics mode compatible printers. The printout will include everything
currently on the screen.
[F6] Save data Save waveform with current settings
[F7] Open data Open waveform with settings
[F8] Protocol
[F9] About us… Short information about OPTEL
[F10] Exit Exit to system
|
| On
the scope screen: |
Filtr - active Toggle on/off
filtering of signal
Pulser[V] - Amplutide of Pulse
Gain[dB] - -20 to 40dB with 1dB step
Delay - Begin of signal window
Range - Length of signal widnow
Offset - Voltage offset
Find First Echo - Search the first echo
<< >> - Search the next signal or the last before the signal
that is displayed
Velocity [m/s]
|

|
Choose
units axis Y between [dB], [%], [V] |

|
Choose
between RF Signal and Detector (absolut, Positiv and Negative) display in
upper window. |

|
Choose
units axis X between [us], [sample], [mm] |

|
Choose
mode of Fourier spectrum display, marker mode allows to choose the signal
region, form which FFT is made |

|
TT
- Through Transmission Mode
PE - Pulse Echo Mode |
| Guide
to control amplify and filter for Special External Pulser/Receiver
with Amplifier |

|
Gain
setting with 0 - 255 levels,corresponding to range 0 - 50 dB
Toggle filter of 0.5 - 15 MHz with 0.5 - 8 MHz |
|
Measuring - Cells: |
Time of flight [us] Time of flight
Thickness [mm] Thickness
|
|
Description of software for measurement of time of flight. |
This software package allows to measure
time of flight.
For each measurement it is necessary to choose reference signal and
compare it with the signal, coming from the measured medium (reflected
or transmitted through it). This allows to use this software with almost
any kind of samples, containments etc. For people using this software it
is necessary to have some knowledge about such kind of measurements,
physics of ultrasounds etc
|
|
The time of flight and thickness measurement method. |
Time of flight - Using white marker in the
upper window the interesing part of the signal should be chosen (or
signal from the first wall).
Thickness - Using blue marker in the upper window the signal from the
second wall should be choosen.
|
|

Guide to OPKUD and OPBOX.pdf
|