Ultrasonic Testing Box
OPBOX- USB 2.0
OPBOX 2.0 Ultrasonic Box with Integrated Pulser and Receiver |
|
|
|
OPBOX ver 1.2 USB 2.0 is a complete ultrasonic testing device, suitable for all ultrasonic measurements and due to many additional digital inputs/outputs and internal processor it can be used as a controller for more complicated
devices.
|
The box can work together with following devices:
- multi channel multiplexer
- Phased Array systems
- Scanner - it has input for
incremental encoders;
The box has an implemented one channel pulser & receiver and can be used with one transducer or with two (one is
sending, and the second one receiving).
Size: 164mm x 138mm x 34 mm
Power Supply: 9V DC / 2A
Weight: 590g
|

|
|
Features: |
- Low noise preamplifier
- Switchable analog filters
- Support fast TGC (time gain compensation)
- 2 signal inputs PE and TT
- Input attenuator (-24dB)
- Input Impedance 50 Ohm, 10pF
- 100MHz sampling frequency
- Delay settings up to 255us
- Various triggering mode
- Interfaces USB 2.0 HiSpeed
- Dimensions: 164mm x 138mm x 34 mm
- BNC or Lemo analog connectors
- DB15 connector for I/O lines
|
|
Technical data |
| A/D
converter: |
| -
Resolution: |
8 bits (internal
10 bits) |
| - Sampling
frequency: |
12.5, 25, 50 or
100MHz 1 |
| Delay
time: |
Post
trigger 255 us1 |
| Data
buffer: |
up to
128kB 1 |
| Analog
parameters: |
| -
Input amplifier gain: |
from
-28dB to 68dB |
| -
Input preamplifier |
0dB
or +24dB |
| -
Sensitivity: |
+/-
0.275V or 2Vpp |
| -
Bandwidth: |
0.5
MHz - 25 MHz (-3dB) |
| -
Input impedance |
50Ohm,
10pF |
| -
Filters switchable: |
0.5 -
6MHz1, 0.5 - 10MHz1, 0.5 - 15MHz1, 0.5 -
25MHz1, 1 - 6MHz1, 1 - 10MHz1, 1 - 15MHz1,
1 - 25MHz1, 2 - 6MHz1, 2 - 10MHz1, 2 -
15MHz1, 2 - 25MHz1, 4 - 6MHz1, 4 - 10MHz1,
4 - 15MHz1, 4 - 25MHz1
|
| DAC (TGC) with arbitrary
waveform generator |
| -
Resolution of time step |
10ns |
| -
Resolution |
8bit |
| -
Max. Gain changing |
48dB
pro step |
| Counters.Input
for incremental encoder: |
2
channels 8bit |
| Trigger |
| - Internal
trigger rate |
software |
| -
External trigger rate |
TTL
Signal |
| -
Multiplexer |
TTL |
| -
Trigger output signal |
TTL
Signal |
| Additional
inputs: |
| - 2
digital lines |
for
external trigger feature |
| - 2
digital lines |
TTL
Signal |
| Output: |
| - 6
digital lines |
TTL
Signal (Multiplexer control up to 64 channels) |
| DATA
BUS |
USB
2.0 (48MHz bulk transfer) |
| Pulser |
| - Pulse
Voltage |
off -
360V (positive pulse, Short circuit step pulser)in 13 levels |
| -
Pulse lenght (loading time) |
from
0 - 3.1us with resolution 0.1us |
| -
Fall Time |
<=20ns |
| -
Pulse duration |
Short
circuit, bandwidth up to about 50MHz |
| -
Output impedance |
<
1Ohm |
| Characteristic
of the pulser on board: |
Pulser circuit
waveforms: The rising edge of the Trigger signal (described as Trigger)
initiates the transducer charging process. After this time (form 0.1 to
3.1us) a transistor switch which discharges the transducer is turned on
(discharge time is about 20ns, but it can be longer, if the capacitance
of the transducer is too large; the limit is reached, if the transducer
is made from the standard ceramics, is 0.1 mm thick and has 8 mm
diameter).
Because of the very low output impedance of the device (<1 Ohm) and
short discharge time the pulse generated with this device could be
concerned as a real pulse answer for the most transducers. Transducers
with a parallel matching inductance should not be used, since they do
not allow the transducer to be pre-charged (the inductance causes a
short-circuit).

|
| Software
Package |
The OPBOX - USB 2.0
software (for Windows 95,98/NT 4.0/2000/XP/Vista/7) includes all necessary
functions for ultrasonic measurements.
|
|

Technical description of the OPBOX device
|