Przedsiębiorstwo Badawczo-Produkcyjne OPTEL Spółka z o.o. 
 ul. Morelowskiego 30 PL-52-429 Wrocław
 phone: +48 71 329 68 53 fax: +48 71 329 68 52 NIP: 898-10-47-033 
 Research & Development / Ultrasonic Technology / Fingerprint recognition

    

OPCARD - 01/100
Ultrasonic board with integrated pulser and receiver - 
Single Channel PCI-bus UT PC-Card

  
 

OPCARD -01/100 is a complete ultrasonic testing device, suitable for all ultrasonic measurements, and due to many additional digital inputs/outputs and internal processor it can be used as a controller for more complicated devices.

The card can work together with following devices:

· 32 channel multiplexer - the card can control it directly;
· Scanner - it has input for incremental encoders;

The card has an implemented one channel pulser & receiver and can be used with one transducer or with two (one is sending and the second receiving).

OPCARD -01/100 is a short PCI card and can be installed in a standard PC.

  Technical data
  A/D converter:
  - Resolution:  8 bit
  - Sampling frequency:  12.5,25,50 or 100MHz1
 Data buffor  256, 512, 1K, 2K,4K,8K,16K, 32k, 64k, 128k1
 Analog parameters:
  - Input amplifier gain:  from -20dB to 76dB1
  - Input attenuator:   off or -20dB1
  - Sensitivity:  0.1mV - 1Vpp
  - Bandwidth:   0.5 MHz - 25 MHz (-3dB)
  - Input impedance  50 Ohm, 10pF
  - Filters switchable   0.5 - 6MHz1, 0.5 - 10MHz1, 0.5 - 15MHz1, 0.5 - 25MHz1
 1 - 6MHz1, 1 - 10MHz1, 1 - 15MHz1, 1 - 25MHz1
 2 - 6MHz1, 2 - 10MHz1, 2 - 15MHz1, 2 - 25MHz1
 4 - 6MHz1, 4 - 10MHz1, 4 - 15MHz1, 4 - 25MHz1
 Pulser:
  - Pulse Voltage  off(0V) to 360V (positive pulse, Short circuit step pulser)
  - Fall Time  <= 20 ns
  - Pulse Duration   Short circuit, bandwidth up to about 50MHz
  - Output impedance  < 1 Ohm
 Delay time:  Post trigger 256us, measurement accuracy  better than 1ns
 Data buffer:  256, 512, 1K, 2K,4K,8K,16K, 32k, 64k, 128k1
 DAC (TGC) with arbitrary waveform generator
  - Frequency  50MHz
  - Resolution  8 bit
  - Max. Gain changing  48dB pro step
 Counters / Input for incremental encoder:  
  - Counters for Incremental Encoders 2 channel  16-bit
 Trigger
  - Internal trigger rate  Software
  - External trigger rate  TTL Signal
  - Multiplexer control  Software
  - Trigger output signal  TTL Signal
 Additional inputs:  
  - 2 x digital line  TTL Signal
 Output:  
  - 5 x digital line  TTL Signal
 Data bus:  PCI (short card)
 Transfer:  Bus Mastering DMA data transfer
   
 

1 - software selected

  Signals on the external connectors


Lemo:
PE - Measured input signal (sending / receiving);
TT - Measured input signal (receiving);

DB15:
Pin1, 9 - TTL encoder input signals
Pin2, 10 - TTL general input signals
Pin4 - TTL trigger out signal
Pin5, 6, 7, 12, 13 - multiplexer channels control lines
Pin8 - +12V
Pin15 - -12V
Pin14 - GND

  Characteristic of the card:


One of the most important features of the card is a precise synchronization between the transmitter trigger signal postTrigger and the moment when the sampling of the input signal starts.

 This time (postTrigger) is software programmable in the range of 0 – 255us with a resolution of 1us and has stability within the range of 1ns. It is particularly important in the case of scanning devices, since it allows to achieve a very small time skew between different positions (channels). 1ns would correspond to a clock frequency of 1GHz which is much higher than the actual frequency used.

 Characteristic of the pulser on board:


Pulser circuit waveforms: The rising edge of the Trigger signal (described as Trigger) initiates the transducer charging process which takes about 3us. After this time a transistor switch which discharges the transducer is turned on (discharge time is about 20ns, but it can be longer, if the capacitance of the transducer is too large; the limit is reached, if the transducer is made from the standard ceramics, is 0.1 mm thick and has 8 mm diameter).

Because of the very low output impedance of the device (<1 Ohm) and short discharge time the pulse generated with this device could be concerned as a real pulse answer for the most transducers. Transducers with a parallel matching inductance should not be used, since they do not allow the transducer to be pre-charged (the inductance causes a short-circuit).

 

 Software Package 


The software (for Windows 95,98/NT 4.0/2000/XP) includes all necessary functions for ultrasonic measurements.