 Research & Development / Ultrasonic Technology / Fingerprint recognition

    


Ultrasonic power generator
MARP version 1.3

  

 Generator MARP ver. 1.3


This generator was created as a part of a device for ultrasonic therapy.
It is designed in modular form and composed from following boards: control board, power generator board and radiator with screening function. All together in sandwich form.

 Technical data
  
  • Working frequency: 1 and 3 MHz 
  • Full digital frequency synthesis, integrated microcontroller. 
  • Maximal power output: 30W; 
  • Two switchable transducer outputs 
  • Impedanc matching for two frequency ultrasonic heads (1 and 3MHz) 1cm2, 4 - 5cm2 and 10 - 12cm2; 
  • Continuous power mode
  • Burst mode with adjusable repetition rate of bursts between 10 and 150Hz; aith 10Hz Step. 
  • Step wise controlled ratio between pause and power: 5%, 10%, 20%, 25%, 50%, 100%; with 5% step. 
  • Power regulated between 5 and 100%; 
  • Analyser of impedance for controlling the contact between transducer and skin.
  • Galvanic isolated output transformer, allowing to apply additional voltage to the electrodes of transducer.
  • All parameters controlled via digial lines with 3,3 V standard.
 Ultrasonic heads

 

Transducer, intended to work with the generator
Two resonance frequencies: 1 and 3 MHz
Active surface: 1cm2, 5cm2, 12cm2
Biocompatible housing


  