Description of a device
and software for precise sound velocity measurement |
The device is used in some chemical
laboratories around the world. It allows very quick and simple
measurement of fluid parameters - among all the concentration of
solutions and monitoring of chemical reactions. The device is sold in
many versions: with ISA-card (as described below), with PCI-card, that
is shown on the picture and with OPBOX USB 2.0
|
| Short
description of the system |
The device contains ultrasonic testing
card OPKUD01/100, pulser&receiver OPGUD and cuvette with transducers
attached on its opposite sides. OPKUD card is controlled by the software,
that makes measurements, communicates with the user and calculate the
sound speed.

System version 1.2

|
| Technical
data - depending on version |
I. Ultrasonic card
- resolution: 8 bit
- sampling rate: 50 or 100MHz1
- channels: 1
- Input voltage: max. 1Vpp
- input impedance: 50Ohm, 10pF
- bandwidth: 0.1-25MHz
- amplifier: -20dB until 40dB
- buffer length: 1-16kB1
- internal (with software)
- external (TTL signal max. 2 kHz)
1 - software
controlled
|
| II.
Pulser&Receiver |
The device works on following basis:
Transducer is used as capacitor, pulser generates a voltage on this
capacitor and than, after a preset level is reached, very quick switch
is used to short circuit the transducer (in about 20ns).
- voltage 50V- 360V (software
controlled, 8 levels possible)
- amplification 40dB
|
| Method
of measurement: |
Cuvette is filled with measured fluid and the time is measured that
sound needs to travel between the walls (in the fluid) and in the walls.
The first or some of the following reflections can be used as result,
that will be evaluated by the software.
A following method for measurement of time of flight of sound is used:
We assume, that the goal of the user is the measurement of fluids, that
has only small differences of sound velocity, and the it is possible to
assume, that a signal practically do not change, only its time of flight.
We are using the feature of the card, that causes, that pulse generated
by the pulser is synchronized with the clock of the card with high
precision (significantly higher than 1ns) and measure the change of
signal amplitude. This allows to achieve highest possible resolution and
accuracy.
It should be told, that it would be also possible to measure another
fluids too, having different acoustical properties (for example larger
attenuation) and thus changing the form of the pulse in the way, that
the above described method will be not suitable. For this purpose
another method can be used - but the precision of time of flight
measurement will be lower.
After the TOF will be estimated, using one or another procedure,
following formulas are used for calculation of sound speed:
We have a fluid container, where sound travels only partially in the
fluid and we can write following equation:

Where:
-
sound propagation time not in the fluid (in walls);
- sound propagation time in fluid.
We can measure propagation
time in the whole system, filled with water
, that has sound velocity ,
or with measured fluid
(sound velocity
). If we know the length of the sound path (
) in measured fluid, we are able to calculate the sound speed in this
fluid:


can be obtained after measurement with water. This measurement should be
done from time to time, but it can be assumed, that it will be not
necessary to repeat it very often, because it is realistic, that
parameters of the system do not change very quickly.

where:
- sound velocity in measured medium
Formula 1. Sound velocity in measured medium
The user of the software (and device) must
know and input in the proper place the length of the sound path in water
( ).
Alternatively it can also input the sound speed (for example in water)
that causes, that software can calculate the path length.
To make a measurement, it is necessary to decide, which signals should
be used for measurement. It is possible to use signals coming after one
time travel through the cuvette or after they was reflected many times.
Signals can be chosen using cursor gates.
It is necessary to tell, that using multiple reflections causes, that
absolute accuracy is lower (it depends on precision of length or sound
speed measurement in water multiplied by the amount of reflections), but
sensitivity (relative accuracy) higher. Thus: choosing multiple
reflections for measurement can be interesting for comparison of
different fluids.
|
| Requirements
for using the software: |
System:
Computer: PC
Graphic card: SVGA with 640x480 or higher resolution (800x600 is
optimum)
System: Windows 95, Windows 98, ME, XP or Windows NT, 2000, 2003
The software is delivered in the version ready for installation. It can
be installed in any directory, chosen by the user. We recommend to
install it in the directory: "C:\OPTEL\". The software creates
two subdirectories: "PROTOCOL" and "SET".
|
| Screens
and functions keys description |
|
Main screen of the software is as follow:

Picture 1. Main screen |
|
Explanation of function of windows and function keys: |
Windows in the left part
of the screen, having to do with a graph screen:
1. The switch in the upper
left part of the graph screen allows to choose units for the vertical
axis. Possible are following units: V, % and db.
2. The switch in the upper right side of the screen allows to set the
method of data visualization. It is possible to show data in the form of
an absolute value (Abs), RF data (HF), only positive value (positive) or
only negative value (neg).
3. The switch in the lower right side of the graph allows to choose
units for the horizontal axis of the graph screen, possible are
following units: ms, mm and sample.
4. Window "Time of flight" shows the time, when the signal in
the first gate arrives.
5. Window "Velocity" shows the calculated sound speed in
measured material.
6. Window "Path Length" shows the calculated sound path length
in measured medium (cuvette length).
Making a mark at the key "Velocity" causes, that the software
calculates the sound speed in measured medium, assuming, that the path
length is known. Making the mark at "Path Length" causes the
calculation of path length.
If multiple reflections are used, it is necessary to set, which
reflection is used in the window "reflection multiplicator".
If the amount of reflection is not correct, the worth showed will be
multiplied by the factor that will depend on the proportion of the
correct worth and chosen one.
Windows on the right side of the screen:
Upper part of the right side of the screen allows to set parameters of
ultrasonic card and pulser&receiver. To chose correct parameters,
some knowledge of the function of card is necessary. The user can also
set parameters, see the results and find the best working ones. The best
setting will occur, if the signal on the screen will have the amplitude
of about 2/3 of screen height.
7. Window "Pulser" allows to set the voltage of the pulser (60
- 360V)
8. "Gain" sets the amplifier on the card.
9. "Delay" defines the delay after trigger, that decides, when
the graph should begin - in relation to the trigger signal.
10. "Range" defines the length of the signal, measured by the
card (length of the signal window).
11. "Offset" allows to shift the signal graph vertically. It
is a software function (not implemented in the hardware).
12. "Average" allows to set the amount of signals, that will
be taken to obtain the signal, that will be used for measurement.
13. Switch "Single/Dual" allows to set, if the software should
work with one transducer, that sends and receives the signal (mode
"single") or with two, where one is sending, and the second is
receiving (mode "dual").
14. The key "Find first echo" starts a function, that seeks
the first echo, and automatically places it on the beginning of the
graph.
15. The keys below allow to find subsequent
16. Beneath the windows can be found, that shows the position of gates
and signal value in this gates. Marking the gate causes, that it can be
switched on or off. Positioning of gate can be done with the mouse and
cursor.
|
|
Description of function keys |
F1 Start/Stop - Starts or stops the
measurement.
F3 Load Settings - Loads stored settings:

F4 Save Settings - Stores
settings chosen by the user:

F5 Print screen - prints
the screen
F6 Save data - Saves data in the file
F7 Load data - Reads data from the file
F8 Protocol - Allows to work
with protocol, loading and storing of protocol, adding data to protocol.

F10 Exit - Exits software
|
| How
to work with the software |
The software starts in the
configuration (parameter settings), that was stored by the user, during
the last session. Before starting the measurements, the user should
input necessary values of parameters (sound speed in water or path
length).
The proper setting of such parameters as gain, depth, level requires
some knowledge, the main rule is: the signal should be seen in the
center of measuring window and his amplitude should be not larger than
2/3 of the window height (and not more than 1Vpp).
After settings are done, the velocity measurement should be started. To
do this, the key [F1] START/STOP should be pressed.
The result of measurement is shown in the following window:

and after it is calculated,
based on the formula 1 in the following fields:

and

|
| Measurement
protocol |
The software allows to make protocol of
measurements. F8 key on the main panel allows to control this. After
each start of the software in the subdirectory protocol a text file with
the name "protokol.txt" is created.
The function "add to file" automatically adds the actual
measurement results to the protocol file, together with description and
date and time of measurement.
In a short time we will add the function, that allows to make
measurements with a given period of time, and store them in a protocol
file.
|
|

|
|
toflen.pdf
|
|